End float test of inhaler cap

  • Mecmesin

Specification

‘End float’ is the diplacement in mm of a product, in this case the cap of the inhaler, measured at a specific load.
The customer wished to apply a 3N load in both tension and compression to the cap of a new design of inhaler. Any resultant displacement from the nominal position then had to be measured.

Solution

The sample was laid horizontally into a slide-mounted nest and clamped into position. The cap of the bottle was rested in a location fixture connected to a 5N S-beam loadcell.

The nest was incremented forward using micrometer adjustment until a minimal compressive load was achieved, displacement at this load was zeroed.

The nest was then moved forward until a 3N alarm point was reached and the displacement of the load was noted (compressive end float). For the calculation of the tensile end float, the process was reversed.


The system was supplied with mechanical limit stops to prevent accidental overload of the small capacity loadcell. The use of both tension and compression visual and audible alarms provided a clear indication of when the test load had been achieved.

System

  • AFTI
  • S-Beam Junior Loadcell 5N
  • Slide Mounted Rest
  • Location Fixture
  • Micrometer Adjustment Barrel
  • Base Plate

Supplied to

GlaxoSmithKline, UK

See also

Tensile testing wire joining systems

Our customer's wire joiners and tensioners are used extensively in the Building Services, Agriculture, Civil Engineering, Viticulture and Seismic Construction sectors and feature award-winning solutions for clamping, seismic bracing, suspension and hanging systems. As part of the continuous improvement ethic, the Quality and Design Department wanted to review their in-house methods and re-evaluate the test procedures for measuring break loads in their cable components.

  • Jeff West
  • Product Info
  • English
  • Created 12 Oct 2016
  • Modified 13 Oct 2016
  • Hits 1304